Charged Aerosol Detectors
Corona Veo and Corona Veo RS
Page 96 Operating Instructions
Warning Code Warning Message Description and Remedial Action
15014 Evaporation heater
fault
The evaporation heater has failed to apply the selected
temperature.
Turn the detector off and on. Test the evaporation heater by
the self test. Allow the detector to come to thermal
stabilization. If the heater still fails, contact Service.
Error Codes
The following table lists the detector-related error codes and messages along with
recommended remedial actions. Error codes immediately result in the interruption of data
acquisition from the detector and will terminate any automates sequences or batch runs when
you operate the detector from a chromatography data system.
If left unattended, error codes may result in permanent physical damage to the detector.
Note the error code or the exact wording of the message and contact the Thermo Fisher
Scientific Service for Dionex HPLC products if you are unable to eliminate the problem.
Error Code Error Message Description and Remedial Action
29006 Low gas pressure The gas pressure has dropped below 241 kPa (35 psig). This can be
caused by an incorrect gas pressure setting: a gas leak (internal or
external), or an incorrect gas supply pressure.
The contact closure was sent to the Digital I/O terminal of the
detector. If a pump is connected to the Digital I/O, the pump flow
will be stopped. On the detector display, press any soft key to
continue.
Corona Veo RS only: The stream-switching valve will divert liquid
flow away from the nebulizer to prevent detector flooding.
Check that
• gas supply pressure is in the range of 480 kPa – 550 kPa
(70 – 80 psig) and that gas is still being delivered to the
detector.
• gas pressure reading on the detector display is the range of
276 kPa – 448 kPa (40 – 65 psig).
• gas supply tubing is correctly installed at the rear of the
detector and that there are no signs of a gas leak.
• gas filters are correctly installed and that there is no sign of
gas leakage.
Turn the detector off and on and allow the self test to test the
pressure sensors. If the error persists, contact Service.